Maintenance Tips
DAILY
-
Clean all vacuum lines and solids
collectors in each operatory.
-
Replenish developer and fixer solution in
X-Ray processor or manual tank.
-
Lubricate handpieces. Remove burs, clean
and lubricate prophy angles and contra angles.
-
Check solution or water in sterilizer.
-
Turn off nitrous oxide and oxygen tanks.
-
Turn off vacuum and compressor.
-
Clean nitrous oxide nasal inhaler hoods
after every use (do not use alcohol).
-
Make sure tubing of cavitron is not
kinked.
-
Turn off main water valves (solenoid) at
night to prevent possible flooding.
WEEKLY
-
Clean X-Ray processor transports.
-
Replaced ultrasonic cleaner solutions.
-
Clean main suction canister.
-
Check oil level in compressor.
-
Check air dryer for desiccant.
-
Clean sterilizer according to
manufacturer's instructions.
-
Check compressor tank for moisture.
-
Clean dental light (make sure it is cool
before touching).
-
Alternate suction pumps (if applicable).
MONTHLY
-
Check nitrous oxide safety level
indicators.
-
Check nitrous oxide unit, lines, hoses,
and masks for leakage.
-
Check air pressure to all handpieces.
-
Clean panoramic cassettes with
intensifying screen cleaner.
-
Check sterilizer with spore test
indicator.
-
Check master water filter element.
-
Clean plaster trap.
SEMI-ANNUALLY
-
Replace the coalescing & intake filters on
air compressor.
-
Lube all moving parts of dental lights &
dental chairs.
-
Check mercury vapour safety levels.
ANNUALLY
|